Getting Started: Installation
1–12
WFM 90 & WFM 91 User Manual
Electrical Installation
Power Source
The WFM 90 and WFM 91 are designed to operate from either six
C-cell batteries, alkaline or rechargeable NiCad, or from an AC
adapter wall unit producing 12 VDC.
CAUTION. DC Power Source
To avoid possible damage to the instrument circuitry when using a
DC power source other than the supplied AC adapter, ensure that the
DC source is a negative-ground 11 – 18 V system with a negative
center lead. The DC source should be able to provide 12 W of power.
Battery Operation
The rechargeable NiCad batteries are automatically recharged when
the 12 VDC adapter is plugged into the instrument. The batteries are
recharged while instrument power is on or off. The battery charge
time is 6.5 hours when the instrument is off, and 20 hours while the
instrument is on. The WFM 90 and WFM 91 shifts into a safe
trickle-charge mode, that can run indefinitely, when the NiCad
batteries are fully charged.
The message
LOW BAT is displayed on-screen when the remaining
battery power cannot guarantee reliable instrument operation. The
length of time the WFM 90 and WFM 91 will continue to operate
after this message is displayed is dependent on the current operating
mode of the instrument.
For optimal battery life and capacity, use the rechargeable NiCad
battery pack in full charge/discharge cycles. In other words, fully
discharge the battery pack before recharging, and then charge the
battery pack until it is fully charged. A new battery pack will take a
few charge/discharge cycles to reach full capacity.